ETHERNET GALVANIC ISOLATOR
ETHERNET GALVANIC ISOLATOR
Inquiry
The Ethernet isolator complies with IEC 60601-1 and EMC/IEC 60601-1-2 safety standards for a network connection between a medical device and an Ethernet network or non-medical device (PC or laptop). In accordance with the IEC EN 60601-1 standard for medical electrical devices and systems, the Ethernet Network Isolator protects the network interface by two independent protective measures (2 MOPP) to the patient and the network connection is safely galvanically isolated in patient care vicinities from dangerous leakage currents that can travel through a copper-based Ethernet network. It effectively isolates medical equipment to protect patients, yet allows data to flow at Gigabit Ethernet speeds. Plugand-play design provides the fast and easy way of installation.
Ordering Information
Technical Data | Property | Property | Property | Property |
Product Number | NIBAe00-01 | NIBBe00-01 | NIBAe00 | NIBBe00 |
Input Interface |
RJ45 Jack, straight | RJ45 Jack, straight | RJ45 Jack, straight | RJ45 Jack, straight |
Output Interface | RJ45 Jack, straight | RJ45 Plug, cable | RJ45 Jack, straight | RJ45 Plug, cable |
Protection Rating | IP40 | IP40 | IP40 | IP40 |
Mating Cycles | > 750 cycles | > 750 cycles | > 750 cycles | > 750 cycles |
Performance Category | C6/Class EA | C6/Class E | C6A/Class EA | C6A/Class EA |
AC Dielectric Strength | 6.0 kV | 6kV | 6.0 kV | 6kV |
DC Dielectric Strength | 8.5 kV | 8.5 kV | 8.5 kV | 8.5 kV |
Total Leakage Current | Typical: 12.0 μA Maximum: 16.0 μA |
Typical: 12.0 μA Maximum: 16.0 μA |
Typical: 12.0 μA Maximum: 16.0 μA |
Typical: 12.0 μA Maximum: 16.0 μA |
Working Voltage | 3000 Vac | 3000 Vac | 3000 Vac | 3000 Vac |
Insulation Section | 2Mopp | 2Mopp | 2Mopp | 2Mopp |
Temperature | -10°C ~ 70°C | -10°C ~ 70°C | -10°C ~ 70°C | -10°C ~ 70°C |
Air Humidity | 10% ~ 90% | 10% ~ 90% | 10% ~ 90% | 10% ~ 90% |
Features
- Isolates harmful leakage currents on a data network to protect patients and equipment
- Easy handling - Plug and play; does not require software, drivers or a power supply
- Compact and universally applicable
- DIN rail mounting compatible by optional bracket
- IEC 60601-1 & IEC 60601-1-2 certified
- IEC 62368-1: 2014/COR1: 2015 and EN 62368-1: 2014/A11: 2017 certified
- UL Listed