The Ethernet isolator complies with IEC 60601-1 and EMC/IEC 60601-1-2 safety standards for a network connection between a medical device and an Ethernet network or non-medical device (PC or laptop). In accordance with the IEC EN 60601-1 standard for medical electrical devices and systems, the Ethernet Network Isolator protects the network interface by two independent protective measures (2 MOPP) to the patient and the network connection is safely galvanically isolated in patient care vicinities from dangerous leakage currents that can travel through a copper-based Ethernet network. It effectively isolates medical equipment to protect patients, yet allows data to flow at Gigabit Ethernet speeds. Plugand-play design provides the fast and easy way of installation.
Technical Data Property Property Property Property
Product Number NIBAe00-01 NIBBe00-01  NIBAe00 NIBBe00
Input Interface
RJ45 Jack, straight RJ45 Jack, straight  RJ45 Jack, straight RJ45 Jack, straight
Output Interface  RJ45 Jack, straight RJ45 Plug, cable  RJ45 Jack, straight RJ45 Plug, cable
Protection Rating  IP40  IP40  IP40  IP40
Mating Cycles  > 750 cycles  > 750 cycles  > 750 cycles  > 750 cycles
Performance Category C6/Class EA  C6/Class E  C6A/Class EA  C6A/Class EA 
AC Dielectric Strength  6.0 kV 6kV  6.0 kV 6kV
DC Dielectric Strength  8.5 kV 8.5 kV  8.5 kV 8.5 kV
Total Leakage Current Typical: 12.0 μA
Maximum: 16.0 μA
Typical: 12.0 μA
Maximum: 16.0 μA
Typical: 12.0 μA
Maximum: 16.0 μA
Typical: 12.0 μA
Maximum: 16.0 μA
Working Voltage  3000 Vac  3000 Vac  3000 Vac  3000 Vac
Insulation Section  2Mopp  2Mopp  2Mopp  2Mopp
Temperature  -10°C ~ 70°C  -10°C ~ 70°C  -10°C ~ 70°C  -10°C ~ 70°C
Air Humidity 10% ~ 90% 10% ~ 90% 10% ~ 90% 10% ~ 90%
  • Isolates harmful leakage currents on a data network to protect patients and equipment
  • Easy handling - Plug and play; does not require software, drivers or a power supply
  • Compact and universally applicable
  • DIN rail mounting compatible by optional bracket
  • IEC 60601-1 & IEC 60601-1-2 certified
  • IEC 62368-1: 2014/COR1: 2015 and EN 62368-1: 2014/A11: 2017 certified
  • UL Listed